Products

Data Logger TDS-602

TDS-602The TDS-602 Data Logger was developed to automatically switch multiple-channel measurements like strain gauge, strain gauge type transducers, DC voltage from a strain gauge and temperature from a thermocouple or platinum resistance temperature device. It has built-in capability for measuring up to 30 points, and can measure up to 1,000 points simply by adding an external switching box. The TDS-602 Data Logger also has a color LCD touch panel that uses a dialog box-based application for quick and simple setting operations. This model has an A/D converter that uses our patented triple integration method with its long field-proven record, and is capable of accurately measuring one channel in 0.06 seconds using regular mode or one channel in 0.02 seconds using high-speed ADC mode. The TDS-602 Data Logger enables measuring points to be set regardless of program settings using manual startup as well as optional LAN connection using a 10-base Ethernet system that facilitates interactive data file transfers to and from computer.
Discontinued


Features

  • Onboard 3.5" floppy disk, large capacity hard disk, high speed printer as recording media
  • Built-in 30 channel switching box (standard 10 channels)
  • High resolution of 0.1×10-6 strain
  • Possible long-term measurement with small-size battery by sleep interval timer
  • Easy touch panel operation on 8.4" color TFT CRT
  • Both GP-IB and RS-232C provided

Specifications

Type TDS-602
Number of channels 1000 channels (using external switching boxes)
Measuring object Strain, DC voltage, thermocouples, Pt RTD
Measuring speed Integral type ADC: 0.06 sec, 0.08 sec/channel
High speed: 0.02 sec/channel
Dimension 430(W)×180(H)×380(D)mm
Weight 13kg
Power supply AC85-250V 50/60Hz 108VA MAX
DC10~15V 12A MAX

Option

  • 10 Base Ethernet mode (Factory installed option)
    LAN
  • Analog output unit (Factory installed option)
    The measured digital values of the channels monitored by TDS-602 are converted into analog voltage values to be output.
  • Alarm Unit ALM-2A
    In the measurement using the alarm function of TDS-602, alarm signals are output.
    Output signals: SEP, UP, LOW, MID