Post-yield Strain Gauge
Post-yield Strain Gauge YEF series
YEF series Post-yield strain gauge
These gauges are applicable to the measurement of large strain up to 10 to 15%. Also these withstand the repeated strain in elastic range (at strain level ±1500 × 10-6 strain) like ordinary strain gauges. However, these are not applicable to the measurement of repeated strain in a large range.
* Suitable when a large strain measurement is made after a few days or more of bonding the strain gauge.
Applicable specimen |
Metal
|
---|---|
Operational temperature(℃)
|
YEF : -30~+80°C YEF: -20~+80°C |
Temperature compensation range(℃)
|
- |
Applicable adhesive
|
CN, CN-Y |
Backing
|
Special plastics
|
Element
|
Cu-Ni |
Strain limit
|
10 to 15 %(100000 to 150000x10-6 strain) |
Fatigue life at room temperature
|
5x105(±1500x10-6 strain) |
-
GOBLET series Single-element / two-element / three-element
Gauge patternTypeGauge
length
(mm)Gauge
width
(mm)Backing
length
(mm)Backing
width
(mm)Resistance
(Ω)Single-element YEFLAB-2 2 1.8 7 4 120 YEFLAB-5 5 1.9 10.5 4 120 90° 2-element Rosette
Plane type
(G.F. 2.1 approx.)YEFCAB-2 2 1.8 9.5 9.5 120 YEFCAB-5 5 2 13.5 13.5 120 45°/90° 3-element Rosette
Plane type
(G.F. 2.1 approx.)YEFRAB-2 2 1.8 9.5 9.5 120 YEFRAB-5 5 2 13.5 13.5 120 * Each package contains 10 gauges.
-
Single-element / two-element / three-element
Gauge patternTypeGauge
length
(mm)Gauge
width
(mm)Backing
length
(mm)Backing
width
(mm)Resistance
(Ω)Single-element YEFLA-2 2 1.8 7.5 4 120 YEFLA-5 5 1.9 12 4 120 90° 2-element Rosette
Plane type
(G.F. 2.1 approx.)YEFCA-2 2 1.8 10 10 120 YEFCA-5 5 2 14.5 14.5 120 45°/90° 3-element Rosette
Plane type
(G.F. 2.1 approx.)YEFRA-2 2 1.8 10 10 120 YEFRA-5 5 2 14.5 14.5 120 * Optional -F availability (lead-free solder)
* Each package contains 10 gauges. -
Leadwire-integrated YEF series (made-to-order)
Application Product name Lead wire code Operational
temperature
rangeLeadwire code
exampledGeneral use (temperature unchanged
during measurement)Paralleled vinyl LJC/LJC-F -20 to +80 ℃YEFLA-5-3LJC General use 3-wire paralleled vinyl LJCT/LJCT-F -20 to +80 ℃YEFLA-5-3LJCT
More Information
Post-yield Strain Gauge YHF series
YHF series Post-yield strain gauge
These gauges are developed for the measurement of very large strain up to 30 to 40%. These are not applicable to the measurement of repeated strain in elastic range as well as in large range.
* Suitable when a large strain measurement is made after a few days or more of bonding the strain gauge.
Applicable specimen
|
Metal
|
---|---|
Operational temperature (℃)
|
-30 to +80 ℃ |
Temperature compensation range (℃)
|
- |
Applicable adhesive
|
CN, CN-Y |
Backing
|
Special plastics
|
Element
|
Special alloy |
Strain limit
|
30 to 40 %(300000 to 400000x10-6 strain) |
Fatigue life at room temperature
|
2x104(±1500x10-6 strain) |
-
Single-element
Gauge patternTypeGauge
length
(mm)Gauge
width
(mm)Backing length (mm)Backing width (mm)Resistance (Ω)Single-element YHFLA-2 2 1.5 8 2.7 120 YHFLA-5 5 1.7 11 3 120 * Optional -F availability (lead-free solder)
* Each package contains 10 gauges. -
Leadwire-integrated YHF series (made-to-order)
Application Product name Lead wire code Operational temperature range Leadwire code
exampledGeneral use (temperature unchanged during measurement) Paralleled vinyl LJC-F -20 to +80 ℃ YHFLA-5-3LJC-F General use 3-wire paralleled vinyl LJCT-F -20 to +80 ℃ YHFLA-5-3LJCT-F
More Information
Post-yield Strain Gauge YF series
YF series Post-yield strain gauge
These gauges are applicable to the measurement of large strain up to 15 to 20%. These are not applicable to the measurement of repeated strain in elastic range as well as in large range.
* Suitable when a large strain measurement is made after a few days or more of bonding the strain gauge.
Applicable specimen
|
Metal
|
---|---|
Operational temperature (℃)
|
-20 to +80 ℃ |
Temperature compensation range (℃)
|
- |
Applicable adhesive
|
CN, CN-Y |
Backing
|
Special plastics
|
Element
|
Cu-Ni
|
Strain limit
|
15 to 20 % (150000 to 200000x10-6 strain) |
Fatigue life at room temperature | - |
-
Single-element
Gauge pattern Type Gauge
length
(mm)Gauge
width
(mm)Backing
length
(mm)Backing
width
(mm)Resistance (Ω) Single-element YFLA-2 2 1.8 7.5 4 120 YFLA-5 5 1.9 12 4 120 YFLA-10 10 2.6 16.6 4.9 120 YFLA-20 20 1.8 26 3.7 120 * Optional -F availability (lead-free solder)
* Each package contains 10 gauges. -
Leadwire-integrated YF series (made-to-order)
Application Product name Lead wire code Operational temperature range Leadwire code
exampledGeneral use (temperature unchanged during measurement) Paralleled vinyl LJC-F -20 to +80 ℃ YFLA-5-3LJC-F General use 3-wire paralleled vinyl LJCT-F -20 to +80 ℃ YFLA-5-3LJCT-F
More Information
Data Logger Use Static Measurement Software Monitoring measurement software Visual LOG Light TDS-700L
Monitoring measurement software Visual LOG Light TDS-700L
Visual LOG® Light is software for Windows for controlling measurement, monitor alarms, etc. using any of our digital static strainmeters.
With its interface and configuration of measurement instruments, this software accommodates direct computer connection by RS-232C / GP-IB / LAN / USB and data memory forwarding, and is optimally designed for each, such that its rich assortment of alarm and monitor functions can still be used immediately on the spot with extreme ease-of-use and a high level of stability.
Type | Applicable measuring instruments | Overview |
---|---|---|
Visual LOG Light TDS-700L |
TS-560 TDS-540(Ver.1.0C or later) TDS-530(Ver.1.0C or later) TDS-150 TDS-302(Ver.1.2A or later) TDS-303 TDS-300 TDS-102(Ver.1.2E or later) TC-32K[single/multi] TC-31K[multi] TC-35N |
Data logger setting, measurement RS-232C/GP-IB/USB/LAN |
-
System configuration
OS / computer / interface
• Computer
OS Microsoft Windows 7(SP1)/8.1/10/11 Computer Any model on which the above OSes can operate, CD drive Interface LAN: 100BASE-TX or better
USB: USB 2.0• Printer
All printers supported by MS-Windows
• GP-IB boar
National Instruments GP-IB board
AT and PC/AT compatibles: AT-GPIB/TNT, AT-GPIB/TNT(PnP), PCMCIA-GPIB*2
NEC PC98 : NEC-GPIB/TNT、NEC-GPIB/TNT(PnP)、PCMCIA-GPIB*2
*1: Windows and MS-Excel are registered trademarks of Microsoft in the US.
*2: If using PCMCIA-GPIB, computer is required to have a PC card slot.
Also check that the PC card service (Windows) is set up.
More Information
Arrester Thunder proof for TML-NET NNZ-2A
Thunder proof for TML-NET NNZ-2A
Protects network measuring system TML-NET from induced lighting. In lighting-prone areas, the induced voltage of lighting traveling through the cables of measuring instruments or network modules can damage network drivers and network modules on the network. When this unit is connected at both ends of the extended network line, it automatically separates the network line when instruments enter standby, to block induced current on the line from lighting and protect networked drivers and modules.
- Features
- Shuts off network line automatically during measurement standby status to avoid induced current
- Power is supplied from network line
- Monitors voltage of network line and current of network modules, and shuts off immediately if abnormal condition occurs
-
Specifications
Type NNZ-2A Surge tolerance 100A (8/20µs impulse) Available number for one system NNZ-2A: up to 10
Low-comsumption network module: up to 100Rated power supply voltage DC18 to 24V Display function Drop of voltage in the network line
Over current of network module (Shuts off the network modules at over current)Environment −20 to +60℃ 85%RH or less (no condensation) Dimensions 50 (W) x 28 (H) x 100 (D) mm (except projecting parts) Weight Approx. 120g
Data Logger Use Static Measurement Software Data Logger use static measurement software TDS-7130v2
Data Logger use static measurement software TDS-7130v2
The TDS-7130v2 runs on a MS Windows PC to perform static measurement using TML data loggers. In addition to visual representation with multiple graphs, this software is a powerful tool for creating a report with data and graphs because it is capable of pasting multiple graphs, value monitors, labels and images on one graph sheet. It can take the most advantage of the data logger function of automatic multi-channel measurement. The TDS-7130v2-A, applicable to long-term measurement and alarm contact, is optionally available. It has additional functions including automatic start of measurement, alarm output using contact unit, and automatic scroll of progress monitor. With these functions, the software is suitable for observation during and after construction, and construction management.
Supported: our data loggers, TS-963, TS-960, TS-560, TDS-540, TDS-630 (Ver.1.1C or later), TDS-530 (Ver.1.0D or later), TDS-300, TDS-302 (Ver.1.2A or later), TDS-303 (Ver.1.0E or later), TDS-601 (Ver.1.4A or later), TDS-602 (Ver.1.0G or later), THS-1000, THS-1100 (Ver.1.1A or later)
- Features
- User interface design suited to wide-screen views
- Supports up to 20 million measurement cycles, continued measurement over a long period
- Interval and data compilator functions create and run 8 tables
- Create all points and 9 data lists, and output data from processing
- Supports TDS-630 fast mode, 0.1-second continuous measurement possible
- Display several graphs on 1 graph sheet
- Vecor/arrow monitor displayable on graph sheet
- Save/recall screen layout
-
System configuration
OS / computer / interface / product key
OS Microsoft Windows 7(SP1)/8.1/10/11 Computer Any meeting specifications recommended by above OSesand having USB port (dual core or better), CD drive Interface LAN 10BASE-T/100BASE-TX
USB USB2.0
RS-232C COM1~COM8
GP-IB NationalInstruments GP-IB
CONTEC GP-IB Interface*
*1 Products with operation verified:
PCI bus supported: GP-IB(PCI)F, GP-IB(PCI)FL
CardBus supported: GP-IB(CB)F, GP-IB(CB)FLProduct key USB dongle Special-priced TDS-7130v2 change-over package available for registered users of TDS/THS-7120 and TOS-7810.
Options
TDS-7130v2-A long-period measurement/alarm contact edition
- External alarm control
Control of the relay control unit to send an alarm outside the computer.
- PHC-D08/PHC-100 PATLITE computer output relay unit
- PHN-R PATLITE network relay control unit (connect up to 3 units)
- Measurement panel auto startup
- Automatic measurement panel display
- Monitor start
- Interval measurement start
- Automatic alarm panel display
- Alarm measurement start
- Alarm interval measurement start
- Automatic text file save
- Create text file of data list when measuring
- Select from any of 10 data lists
- Create text file of occurrence conditions of alarm
- Select file format (text format/CSV format)
- Addition of elapse monitor
Create elapse graph for any designated number of days or time unit.
More Information
Data Logger Use Static Measurement Software
Arrester NZ-6B
Arrester NZ-6B
An arrester protects measuring instruments and transducers from induced lighting. In lighting-prone areas, the induced voltage of lighting traveling through the cables of measuring instruments or switch boxes can damage instruments and transducers. When placing measuring instruments and transducers, connection of NZR-7B/NZ-7C on cabling between an instrument and a switch box or NZ-6B on cabling between a transducer and switch box will protect the instrument and transducer by allowing induced current to pass to the ground.
- Features
- Large discharge capacity
- Equalized discharge circuit
- Fully waterproof
- Applicable to KM transducers with relative temperature measurement function
- Accommodates input common 2-point sensor (2-axis inclinometer)
-
Specification
Type NZ-6B Applicable sensor Strain gauge
Quarter Bridge 3-wire,
Half Bridge,
Full Bridge (including transducer)
Full bridge + Quarter Bridge 3-wire (transducer with rerative temperature measurement function)
Full bridge + Thermocouple
2-axial transducer having common input (2-axial inclinometer)Surge current resistance 10kA (8/20µs) Dimensions ø24mm, Length 225mm Weight 230g